Model NGA2000 NDIR
Instruction Manual
From MAIN MENU
⇓
Technical configuration menu
System set up...
Service menus...
Diagnostic menus...
Other module diagnostic menus...
Listing of all modules...
For Service set up and Service menus, see preceding flow charts.
Diagnostic menus
Control module diagnostics...
Analyzer module diagnostics...
Module Tag
Module Tag
Module Tag
Module Tag
Module Tag
Module Tag
Module Tag
Module Tag
Other module diagnostics
This screen reserved for other modules if present.
Module Tag
Module Tag
Module Tag
Module Tag
Module Tag
Module Tag
Module Tag
Module Tag
Control Unit Diagnostics
See Platform manual.
#8
Analyzer Diagnostics
Power supply voltages...
Primary variable parameters...
Physical measurements...
Temperature control parameters...
Miscellaneous control parameters...
Barometric pressure parameters...
Software diagnostics...
For I/O Module diagnostics, see I/O Modules manual
Analyzer diagnostics
Power supply voltages |
|
+15V analog is: | 14.98V |
+15V analog was: | 14.92V |
+5V digital is: | 5.02V |
+5V digital was: | 4.98V |
+24V power is: | 2.38V |
+24V power was: | 2.42V |
+12V analog is: | 11.88V |
+12V analog was: | 12.0V |
Primary variable parameters
Raw measurement signal: | 0.3258 |
Signal gain setting: | 4% |
Oscillator tune: | 85% |
Chopper speed: | 5 Hz |
Source current: | 1.212 A |
Modulation check... |
|
Percent modulation: | 20% |
Calibration time out: | 60 Secs |
Calibration factors... |
|
0 ppm |
Physical measurements
See #6 in Figure
Temperature control
Fan lower setpoint: | 60°C |
Fan upper setpoint : | 20°C |
Minimum fan duty cycle: | 30% |
Case temperature: | 45°C |
Detector setpoint: | 62°C |
Detector P gain: | 32.52 |
Detector I gain: | 1.34 |
Detector bias: | 283°C |
Detector temperature: | 62°C |
Modulation check
Measurement range number: | 1 | |
Detector signal: |
| 7.0 |
Signal gain setting: |
| 4% |
Status | Ready | |
Instructions: | Flow zero | |
Then: |
| Watt |
Percent modulation: |
| 20% |
(updated only at the end of this test)
Calibration Factors
See #3 in Figure
HISTORY
Primary variable parameters
Signal gain settings: | 4% |
Oscillator tune: | 85% |
Source current: | 1.212 A |
Percent modulation: | 85% |
Listing of all modules |
|
|
Lists all modules detected on the |
|
|
network. |
|
|
Jumps to the module’s diagnostic |
|
|
screen. |
| Analyzer Diagnostics |
Module Tag |
| See #8 above. |
Module Tag |
|
|
Module Tag |
|
|
Module Tag |
|
|
Module Tag |
|
|
Module Tag |
|
|
|
|
|
Miscellaneous control parameters
Fan current: | 35 mA |
Fan duty cycle: | 50% |
Heater current: | 1.7A |
Heater duty cycle: | 50% |
Source current pot setting: | 50% |
Actual source current: | 1.212A |
Alarm messages valid for: | FAILURE |
Barometric pressure parameters
Pressure transducer: | Present |
Barometric pres. Compensation: | Enabled |
Measured pressure: | 12.3 kPa |
Transducer offset: | 1013 hPa |
Transducer slope: | 1.0 |
Transducer PGA gain: | 1.0 |
Software diagnostics
Last message: | No error |
And: | No error |
And: | No error |
And: | No error |
And: | No error |
And: | No error |
And: | No error |
And: | No error |
Edit to reset: | Report |
Software error code (1 = no error): | 1 |
Figure
Rosemount Analytical Inc. A Division of Emerson Process Management | Maintenance and Service |