Model NGA2000 NDIR
Instruction Manual
4
Main Menu
Basic controls...
Expert controls and setup...
(Operational configuration)
Technical level configuration...
(Diagnostic and manufacturing/service)
Basic level security
(expert, technical)
Enter your security code with softkeys. Security codes are set in the system setup menu in the Technical level menus.
Press five characters when Security code says READY. Press six to return to “READY”.
Security code: | 54321 |
Basic controls
Measurement range number: | 1 |
Range upper limit: | 10 ppm |
Range and functional control: LOCAL
Zero gas concentration: | 0 ppm |
Span gas concentration: | 9.89 ppm |
Ranges with valid calibration: | 1&2 |
Calibration status: | Ready |
If it won’t calibrate... |
|
Status: | Ready |
ZERO
|
| Current measurement parameters |
|
|
|
|
| |
|
| Measurement range number: | 1 |
|
|
| Current measurement parameters | |
|
| Range change control: | Local |
|
|
| Response time: | 2.0 Secs |
PARMS |
| Measured Gas: | CO2 |
| MORE |
| Sample flow: | 1000 ml/min. |
|
| Linearization mode: | ON |
|
|
| Sample pressure: | 12.3 kPa |
|
| Gas Measurement Setup: | ON |
|
|
| Case Temperature: | 55.3!C |
|
|
| ||||||
|
| Analyzer operational state: | Normal |
|
|
|
|
|
|
|
|
|
| ||||
|
| Analyzer alarm state: | Normal |
|
|
|
|
|
|
| Alarms report: | Failure |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Expert controls and setup |
| Technical configuration menu |
Expert analyzer controls... |
| System set up... |
Auxiliary module controls... |
| Service menus... |
System set up... |
| Diagnostic menus... |
Analyzer module set up... |
| Other module diagnostic menus... |
Auxiliary module set up... |
| Listing of all modules... |
...Local I/O set up |
|
|
|
|
|
Expert and Technical menus follow in succeeding flow charts.
SPAN
Analyzer Zero
Are you sure?
You must have zero gas flowing through the analyzer.
This control does NOT control any auto- cal. Module bound to this analyzer!
If you are sure, press ZERO again now. Press left arrow key when you are done.
Calibration status:Ready
Error message for last zero:
Analyzer Span
Are you sure?
You must have zero gas flowing through the analyzer.
This control does NOT control any auto- cal. Module bound to this analyzer!
If you are sure, press SPAN again now. Press left arrow key when you are done.
Calibration status:Ready
Error message for last zero:
If it won’t calibrate...
Check that you are flowing the correct gas, and the gas concentration is what it is supposed to be. Make sure that the reading is stable before starting. If you have changed the range fullscale value or any linearizer coefficients, or enabled or disabled it, or done anything else that would affect how it measures the gas, you may have made it hard for the analyzer to calibrate. If so, manually adjust the coefficients until the readings are close to correct, and try again.