Instruction Manual
Analyzer Module Set Up, and
Calibration Parameters.
2.Using the ↓ arrow, select Zero Ranges, press ENTER and, using the up/down arrows, toggle to SEPARATE. Do the same for the Calibrate Ranges selection. Do not press ESCAPE at any time unless retention of prior settings is desired.
3.Return to the Main Menu and make the following selections: Expert Controls and Setup, Expert Controls, CAL DATA softkey, FACTORS softkey, and Range 1 (2, 3, 4) Factors (do Steps 4 and 5 for each range).
4.Select Zero Offset, press ENTER, adjust the value to 500000 with the ↑ and ↓ arrow keys, and press ENTER. Do not press ESCAPE at any time unless retention of prior settings is desired.
5.Refer to the Data Sheet in the rear of this manual for Span Factors as originally set at the factory. Select Span Factor, press ENTER, adjust the value to match the values on the Data Sheet with the ↑ and ↓ arrow keys, and press ENTER. If Data Sheet is not available, enter 0.000015 with the ↑ and ↓ arrow keys, and press ENTER. Do not press ESCAPE unless retention of prior settings is desired.
6.Attempt to recalibrate the Analyzer Module according to the procedure outlined at the beginning of Section 3- 5 on page
Another cause of failure to calibrate is the following: The value for "Maximum range" is lower than the upper limit value for the range in use. See the Range Settings menu for this information.
DISPLAY MESSAGE |
| DESCRIPTION | TYPE |
BAROMETER |
| System Barometer | WARNING |
CASE TEMP |
| Case Temperature | WARNING |
CHOP SPEED |
| Chopper Speed | WARNING |
CRUDE NOISE |
| Calculated Noise | WARNING |
DET SIG |
| Detector Signal | WARNING |
DET TEMP |
| Detector Temperature | WARNING |
LIN ERROR |
| Linearizer Error | WARNING |
N15 VOLTS |
| Power Supply, | WARNING |
P12 VOLTS |
| Power Supply, +12V | WARNING |
P15 VOLTS |
| Power Supply, +15V | WARNING |
P24 VOLTS |
| Power Supply, +24V | WARNING |
P5 VOLTS |
| Power Supply, +5V | WARNING |
PERCENT MOD |
| Percent Modulation | WARNING |
RAW SIGNAL |
| Raw Signal | WARNING |
SVFLOW |
| Sample Bypass Flow | WARNING |
SW ERROR |
| Software Error | FAILURE |
Table | NDIR Analyzer Module Alarms |
|
Rosemount Analytical Inc. A Division of Emerson Process Management | Startup and Operation |