IQS-12001B
Cable Assembly and Component Test Systemwww.EXFO.com
SPECIFICATIONSa
FTB-85102 FTB-8510-12 FTB-8510-2
Ports Gigabit Ethernet Two 10/100Base-T Two 10/100Base-T Two 10/100Base-T and two
and one Gigabit Ethernet
Connector types RJ-45 (ISO 8877) RJ-45 (ISO 8877) and LC RJ-45 (ISO 8877) and LC
Connect speed (Mb/s) 10/100 10/100/1000 10/100/1000
Duplex mode Full/half-duplex Full/half-duplex Full/half-duplex
auto-negotiation auto-negotiation auto-negotiation
Maximum port capacity (Mb/s) 200 (bidirectional) 2000 (bidirectional) 2000 (bidirectional)
Ethernet testing RFC 2544 R FC 2544 RFC 2544
RFC 1242 RFC 1242 RFC 1242
SPECIFICATIONSa
Singlemodeb(1310/1550 nm) Singlemodeb(1490/1625 nm) Multimodec
Testing time d(s) < 6 < 6 < 6
Cable assembly length (m) 1.8 to 15 00 1.8 to 1500 1.8 to 500
Insertion loss measurement uncertainty ± 0.03 e± 0.03 e± 0.07 f
Insertion loss measurement stability g(dB) < 0.005 < 0.005 < 0.028
Reflectance range (dB) −30 to −75 −30 to −75 −10 to −50
Reflectance measurement uncertainty (dB) ±1 (−30 to −70) ±1 (−30 to −70) ± 1.2 (−10 to –30)
± 1.5 (−30 to –40)
±2.2/±1.7 (−70 to −75) ±1.7 (−70 to −75) ± 1.6 (−40 to –43)
± 2.9 (−43 to –50)
Reflectance measurement repeatability h(dB) ±0.1 (−30 to −65) ±0.1 (−30 to −65) ± 0.2 (−10 to –30)
±0.2 (−65 to −70) ±0.2 (−65 to −70) ± 0.4 (−30 to –40)
±0.4 (−70 to −75) ±0.5 (−70 to −75) ± 0.6 (−40 to –43)
± 1.8 (−43 to –50)
Wavelengths (nm) 1310/1550 1490/1625 850/1300
Output channels Up to 32 channels Up to 32 channels Up to 32 channels
Test method End-to-end/bidirectional End-to-end/bidirectional End-to-end/bidirectional
Notes
a. Calculated and measured with the following considerations:
• Based on recommended procedure for ORL measurements
• The FOA-300 series’ uncertainty is included in all uncertainty values
• Although tests are possible with other fiber-optic adapters, these specifications are only valid with the FOA-300 series
• Cable assemblies < 1.8 m can be tested using a non-reflective termination
• With 15 minute warm-up time
b. For SMF-28 fiber. At 23 ºC ± 1 ºC in High-Sensitivity Reflectance mode with no switch.
c. At 850 nm and 1300 nm. For 62.5/125 µm fibers. At 23 °C ± 1 °C, after 30 minutes warmup. Switch module connectorized with FC/PC connectors.
Uncertainty due to launching conditions not included.
d. Calculation and storage time for simplex, dual-wavelength IL/reflectance end-to-end measurements (Standard Reflectance Sensitivity mode, 500 meter range);
does not include referencing and connection time.
e. For simplex measurements using FOA-322 adapter. Uncertainty for MTP to MTP, MT-RJ to MT-RJ, MTP to fanout and MT-RJ to fanout is ± 0.06 dB
(reported with a level of confidence of 95 %). This does not include uncertainties due to connector, connector adapter or switch PDL.
f. For simplex measurement using FOA-322 adapter. Uncertainty for MTP to MTP, MT-RJ to MT-RJ, MTP to fanout and MT-RJ to fanout is ± 0.08 dB
(calculated for a level of confidence of 95 %). Does not include uncertainties due to connector, connector adapter or uncertainties due to mode dependence of the switch.
g. For a stable connection, over 15 minutes, at constant temperature.
h. For a stable connection, over 10 measurements.

GENERAL SPECIFICATIONS

Communication interfaces
General-purpose interface bus (GPIB)
Ethernet (10/100 Base-T)
RS-232
Temperature
Operating 0 °C to 40 °C (32 °F to 104 °F)
Storage −40 °C to 60 °C (−40 °F to 140 °F)
Relative humidity 80 % maximum, non-condensing

Laser Safety

21 CFR 1040.10 CLASS 1 LASER PRO DUCT
IEC 60825-1:2001 CLASS 1 LASER PRODUCT

Standard Accessories

Expansion units as required, user guide, certificate of compliance, custom report, label generator, IQS-12001B software and EUI-89 FC narrow key connector.