Field Controls CC 2000 manual 1000, Control

Models: CC 2000

1 8
Download 8 pages 49.77 Kb
Page 7
Image 7

Table 3: UV-18X Test data and results (testing with lamp on)

Sample

Air

Airflow

Serratia

Corrected

CFU/FT3

%Survival

Log

%

Sampling

Velocity

marcescens

Particle

of air

Number

Duration

Positive hole

CFU/Control

Reduction

Effective

(fpm)

Counts

(count/min)

 

(min)

 

count

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

2.5

1000

21

22

 

 

 

 

2

2.5

1000

27

28

 

 

 

 

 

 

 

 

 

 

 

 

 

3

2.5

1000

28

29

 

 

 

 

 

 

 

 

 

 

 

 

 

4

5

1000

48

51

 

 

 

 

Total min = 12.5

 

Total Corrected

= 130

10.40

9.67

1.01

90.33

 

 

 

Particle Counts

 

 

 

 

 

 

 

 

 

Control:

107.50

 

 

 

 

 

 

 

 

 

 

 

 

1

5

500

8

8

 

 

 

 

2

5

500

10

10

 

 

 

 

 

 

 

 

 

 

 

 

 

3

5

500

17

17

 

 

 

 

 

 

 

 

 

 

 

 

 

4

10

500

28

29

 

 

 

 

Total min = 25

 

Total Corrected

= 64

2.56

0.95

2.02

99.05

 

 

 

Particle Counts

 

 

 

 

 

 

 

 

 

Control:

269.00

 

 

 

 

 

 

 

 

 

 

 

 

Page 7

Page 7
Image 7
Field Controls CC 2000 manual 1000, Control