1523, 1524 Thermometer Readout

Specifications

Table 9 Ohms Measurement, Thermistor

Ohms Range

Accuracy ± Ω, 4 Wire

 

 

200 Ω to 50 kΩ

± (0.01 % + 0.5 )

 

 

50 kΩ to 500 kΩ

± (0.03 %)

 

 

Temperature Coefficient ( –10 °C to 13 °C , +33 °C to 60 °C):

0.002 %/°C + 0.1 (0 Ω to 50 kΩ)

0.06 %/°C + 0.1 (50 kΩ to 500 kΩ)

Excitation Current:

10 A (0 Ω to 50 kΩ)

2 A (50 kΩ to 500 kΩ)

Equivalent temperature accuracies derived from primary specifications (, mV)

Table 10 Temperature, Thermocouples

 

 

 

Measure

Type

 

Range

Accuracies (ITS-90)

 

 

 

 

B

600

°C to 800 °C

0.85 °C

 

800 °C to 1000 °C

0.68 °C

 

1000

°C to 1800 °C

0.57 °C

 

 

 

 

C

100

°C to 550 °C

0.32 °C

 

550 °C to 2300 °C

0.71 °C

 

 

 

E

–200 °C to 0 °C

0.52 °C

 

0 °C to 950 °C

0.22 °C

 

 

 

J

–200 °C to 0 °C

0.52 °C

 

0 °C to 1200 °C

0.23 °C

 

 

 

K

–200 °C to 0 °C

0.61 °C

 

0 °C to 1370 °C

0.24 °C

 

 

 

 

26

Page 34
Image 34
Fluke 1523, 1524 manual Excitation Current