Specifications
Number of Turns | 50 |
Maximum Current | 11A rms, continuous |
| 20.5A rms, 2 minutes |
|
|
Maximum Duty Cycle Derating | < 11A, continuous |
| > 11A, 2 minutes ON, 8 minutes OFF |
|
|
Maximum Voltage | 7 V dc or 7 V rms ac |
Minimum Inner Diameter of Clamp Jaws | 2.54 cm [1”] |
Specification for Clamp Meter/Coil interaction
Specification for 5500A/COIL
|
|
|
|
|
| Coil Specifications [clamp/coil interaction] |
| |||
| Calibrator Output | Effective |
|
|
|
|
|
| ||
|
|
|
| |||||||
| Current Output | |||||||||
|
|
|
| [such as the Fluke | ||||||
|
|
|
|
| [such as the Fluke | |||||
|
|
|
|
| ||||||
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
| ||||
Magnitude | Frequency | ±(% of output + Amps) | ±(% of output + Amps) | |||||||
0.2 | - 0.33 A | DC | 10 - 16.4999 | 0.25 |
| 0.002 | 0.50 |
| 0.02 | |
0.33 - 2.9999 A | DC | 16.5 - 149.999 | 0.25 |
| 0.015 | 0.5 |
| 0.14 | ||
3.0 | - 20.5 A | DC | 150 - 1025 | 0.25 |
| 0.05 | 0.5 |
| 0.5 | |
0.2 | - 0.33 A | 45 | Hz to 65 Hz | 10 - 16.4999 | 0.28 |
| 0.003 | 0.56 |
| 0.03 |
0.33 - 2.9999 A | 45 | Hz to 65 Hz | 16.5 - 149.999 | 0.28 |
| 0.025 | 0.56 |
| 0.25 | |
3.0 | - 20.5 A | 45 | Hz to 65 Hz | 150 - 1025 | 0.28 |
| 0.09 | 0.56 |
| 0.9 |
0.2 | - 0.33 A | 65 | Hz to 440 Hz | 10 - 16.4999 | 0.79 |
| 0.003 | 1.00 |
| 0.03 |
0.33 - 2.9999 A | 65 | Hz to 440 Hz | 16.5 - 149.999 | 0.79 |
| 0.027 | 1.00 |
| 0.25 | |
3.0 | - 20.5** A | 65 | Hz to 440 Hz | 150 - 1025 | 0.79 |
| 0.1 | 1.00 |
| 0.9 |
0.2 | - 20.5 A | Above 440 Hz | 10 - 1025 |
|
| Usable but unspecified |
| |||
|
|
|
|
|
|
|
|
|
|
|
**Note: Some calibrators may not have adequate compliance voltage to drive coil in this range.
Calculating Total Specification
The total specification of the effective current that the
Stotal = S 2 coil + U 2 source
Example:
Assume we are driving the coil with the Fluke 5500A at 4A, 60 Hz (the clamp meter will see an effective current of 200A, 60 Hz) and we are calibrating a
Specification of effective calibrator current in coil bundle = ± (0.06% + 0.1A) = 0.11% Specification due to clamp Meter/Coil Interaction = ± (0.28% + 0.09A) = 0.325%
The RSS of these two specifications determines the total specification of the clamp/source combination:
S total = 0.325% 2 + 0.11% 2 = 0.343%