3 Specifications and Warranty

3.1 Specifications

See Table2.
TemperatureRange –200°Cto 156°C
EntireAssembly Range –200°Cto 156°C
Accuracy(Includes calibration
uncertaintyand short-term
stability)
±0.05°C(full range)
Long-termDrift (°C) <±0.01at 0°C per year, when used periodically at 200
Hysteresis(°C) <0.01
TemperatureCoefficient 0.003925ohms/ohms/°C
Size(O.D.) 0.25"(6.35 mm)
SheathLength 6"(152.4 mm)
SheathMaterial Inconel600
Cablelength 20ft. (609.6 cm)
LeadWires 22AWG/ Teflon insulated cable, 4-wire

3.2 Warranty

Hart Scientific,Inc. (Hart) warrants this product to be free from defects in ma-
terialand workmanship under normal use and service for a period as stated in
our current product catalog from the date of shipment. Thiswarranty extends
only tothe original purchaser and shall not apply to any product which, in
Hart’ssole opinion, has been subject to misuse, alteration, abuse or abnormal
conditionsof operation or handling.
Softwareis warranted to operate in accordance with its programmed instruc-
tionson appropriateHart products. It is not warranted to be error free.
Hart’sobligation under this warranty is limited to repair or replacement of a
product whichis returned to Hart within the warranty period and is determined,
upon examinationby Hart, to be defective. If Hart determines that the defect or
malfunction has been caused by misuse,alteration, abuse or abnormal condi-
tionsor operationor handling,Hart will repair the product and bill the pur-
chaser for the reasonablecost of repair.
5623A 7
3Specifications and Warranty
Table2 Specifications