Fluke 80 Series III user manual MIN MAX Recording Specifications

Models: 80 Series III

1 58
Download 58 pages 10.8 Kb
Page 58
Image 58

80 Series III

Users Manual

Table 19. MIN MAX Recording Specifications

Model

83

85, 87

Nominal Response

100 ms to 80%

1 s

100 ms to 80% (DC functions)

120 ms to 80% (AC functions)

1 s

250μs

(Model 87 only)

Accuracy

Specified accuracy ±12 counts for changes >200 ms in duration (±40 counts in ac with beeper on)

Same as specified accuracy for changes >2 seconds in duration (±40 counts in ac with beeper on)

Specified accuracy ±12 counts for changes >200 ms in duration

Specified accuracy ±40 counts for changes >350 ms and inputs >25% of range

Same as specified accuracy for changes >2 seconds in duration

Specified accuracy ±100 counts for changes >250 μs in duration (± 250 digits typical for mV, 400 μA dc, 40 mA dc, 4000 mA dc)

50