Maintenance
Performance Test
Verify the probe accuracy by measuring a 25 kV dc ±0.25% voltage source. When used with a compatible dc voltmeter, the probe should measure the source with ±1% accuracy. No calibration adjustments are provided.
Cleaning
Use a soft cloth dampened in distilled water to clean the
LIMITED WARRANTY & LIMITATION OF LIABILITY
Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs and services are warranted for 90 days. This warranty extends only to the original buyer or
Fluke authorized resellers shall extend this warranty on new and unused products to
Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center or send the product, with a description of the difficulty, postage and insurance prepaid (FOB Destination), to the nearest Fluke authorized service center. Fluke assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that the failure was caused by misuse, alteration, accident or abnormal condition of operation or handling, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR ANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
Fluke Corporation | Fluke Europe B.V. |
P.O. Box 9090 | P.O. Box 1186 |
Everett, WA | 5602 BD Eindhoven |
U.S.A. | The Netherlands |
5/94 |
|
®
80K-40 High Voltage Probe
Instruction Sheet
Introduction
The Model
Specifications
The
†Voltage Range: 1 kV to 40 kV dc or peak ac, 28 kV rms ac
Input Resistance: 1000 MΩ
Division Ratio: 1000: 1 (1000X attenuator)
Accuracy DC:
20kV to 35 kV: ±1% at 20°C to 30°C; add 1% at 10°C
<20°C and >30°C to 45°C. (For total measurement accuracy add accuracy specification of voltmeter being used.)
0 kV to <20 kV and >35 kV to 40 kV: ±2%. Accuracy AC: 60 Hz, ±5%.
Safety: Meets IEC
*The input impedance of Autoranging Fluke handheld digital
multimeters varies as a function of range. The only range that deviates significantly from 10 MΩ is the 3V (Models 21, 23, 25,
27, 70, 73, 75, 77) or 4V (Models 10, 11, 12, 29, 79, 83, 85, 86, 87, 88) range where the impedance is 11.11 MΩ. To enhance the measurement accuracy when using this range, apply a correction factor of 0.99, i.e. multiply the displayed reading by .99.
†This probe is intended for low energy applications such as CRT and similar circuits. Above 2000 meters altitude, and up to 5000 meters, derate linearly the working voltage from 40 kV peak to 28 kV peak, and derate linearly the transient overvoltage from 80 kV peak to 57 kV peak. Transient overvoltage refers to
Measurement Considerations
Before attempting to use the
PN 481978
January 1978, Rev.9, 3/97
© 1978, 1996, 1997 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies.