48 HartScientific
9Digital Communication Interface
CommandDescription
Command
Format
Command
Example Returned
Returned
Example
Acceptable
Values
Setprogram function to npf=npf=2 1to 4
ConfigurationMenu
ProbeMenu
ReadR0 calibration parameter r[0] r r0:999.999 r0:100.578
SetR0 calibration parameter to nr[0]=n r=100.324 98.0to 104.9
ReadALPHA calibration parameter al[pha] al al:9.9999999 al:0.0038573
SetALPHA calibration parameter to nal[pha]=n al=0.0038433 .00370to .00399
ReadDELTA calibration parameter de[lta] de de:9.99999 de:1.46126
SetDELTA calibration parameter to nde[lta]=nde=1.45 0.0to 2.9
ReadBETA calibration parameter be[ta] be be: 9.999 be:0.342
SetBETA calibration parameter to nbe[ta]=nbe=0.342 –100.0to 100.0
OperatingParameters Menu
Settemperature units: u[nits]=c/f CorF
Settemperature units to Celsius u[nits]=c u=c
Settemperature units to Fahrenheit u[nits]=f u=f
Readcutout mode cm[ode] cm cm:{xxxx} cm:AUTO
Setcutout mode: cm[ode]=r[eset]/a[uto] RESETor AUTO
Setcutout to be reset manually cm[ode]=r[eset] cm=r
Setcutout to be reset automatically cm[ode]=a[uto] cm=a
Readapproach setting ap[proach] ap ap:9 ap:5
Setapproach setting to ndegrees ap[proach]=nap=15 0to 20°C
Readstability ts ts ts:9.9 ts:0.5
Setsoak stability to ndegrees ts=nts=.1 .01to 4.99°C
SerialInterface Menu
Readserial sample setting sa[mple] sa sa: 9 sa:1
Setserial sampling setting to nseconds sa[mple]=nsa=0 0to 4000
Setserial duplex mode: du[plex]=f[ull]/h[alf] FULLor HALF
Setserial duplex mode to full du[plex]=f[ull] du=f
Setserial duplex mode to half du[plex]=h[alf] du=h
Setserial linefeed mode: lf[eed]=on/of[f] ONor OFF
Setserial linefeed mode to on lf[eed]=on lf=on
Setserial linefeed mode to off lf[eed]=of[f] lf=of
CalibrationMenu (WARNING – changing the following calibration values may change the accuracy of the instrument.)
ReadB0 calibration parameter *b0 *b0 b0: 9 b0:0
SetB0 calibration parameter to n*b0=n *b0=0 –999.9to999.9
ReadBG calibration parameter *bg *bg bg: 999.99 bg:15625
SetBG calibration parameter to n*bg=n *bg=156.25 –999.9to999.9
CommunicationsCommands Continued