9105/9107 47
9Digital Communication Interface
CommandDescription
Command
Format
Command
Example Returned
Returned
Example
Acceptable
Values
DisplayTemperature
Readcurrent set-point s[etpoint] s set:9999.99 {C or F} set:150.00 C
Setcurrent set-point to n s[etpoint]=n s=450 InstrumentRange
Readscan function sc[an] sc scan:{ONor OFF} scan:ON
Setscan function:sc[an]=on/of[f] ON or OFF
Turnscan function on sc[an]=on sc=on
Turnscan function off sc[an]=of[f] sc-of
Readscan rate sr[ate] sr srat:999.99 {C or F}/min srat:10.0 C/min
Setscan rate to ndegrees per minute sr[ate]=nsr=5 .1 to 100°C
Readdisplay temperature hold status hm[ode] hm hm:{OFFor AUTO or NO or NC}
Setdisplay temperature hold mode: hm[ode]=OF[F]/AU[TO]/N
O/NC
OFForAUTOorNO
orNC
Sethold mode to off hm[ode]=OF[F] hm=of
Sethold mode to automatic hm[ode]=AU[TO] hm=auto
Sethold mode to normally open hm[ode]=NO hm=no
Sethold mode to normally closed hm[ode]=NC hm=nc
Readtemperature t[emperature] t t: 9999.99 {C or F} t:55.69C
Readhold status ho[ld] ho ho:{Closedor Open}, 9999.99 {C
orF}
ho:Open, 75.0 C
SecondaryMenu
Readproportional band setting pr[op-band] pr pb: 999.9 pb: 15.9
Setproportional band to npr[op-band]=n pr=8.83 Dependson
Configuration
Readcutout setting c[utout] c c:9999 {C or F} c: 620 C, in
Setcutout setting: c[utout]=n/r[eset]
Setcutouttondegrees c[utout]=n c=500 TemperatureRange
Resetcutout now c[utout]=r[eset] c=r
Readheater power
(dutycycle)
po[wer] po p%:9999 po: 1
Rampand Soak Menu
Readnumber of programmable set-points pn pn pn: 9 pn:2
Setnumber of programmable set-points to npn=npn=4 1 to 8
Readprogrammable set-point number npsnps3 psn: 9999.99 {C or F} ps1: 50.00C
Setprogrammable set-point number nto npsn=nps3=50 1to 8, Instrument
Range
Readprogram set-point soak time pt pt ti:999 ti:5
Setprogram set-point soak time to nminutes pt=npt=5 0to 500
Readprogram control mode pc pc prog: {OFF or ON} prog:OFF
Setprogram control mode: pc=g[o]/s[top]/c[ont] GOor STOP or CONT
Startprogram pc=g[o] pc=g
Stopprogram pc=s[top] pc=s
Continueprogram pc=c[ont] pc=c
Readprogram function pf pf pf:9 pf:3
Table2 Communications Commands