Un= F New units selected
7.3 Scan
The scan rate can be set and enabled so that when the set-pointis changed the
dry-well heats or cools at a specified rate (degrees per minute) until it reaches
the new set-point. Withthe scan disabled the dry-well heats or cools at the
maximum possible rate.
7.3.1 Scan Control
The scan is controlled with the scan on/offfunction that appears in the main
menu after the temperature scale units.
Sc=OFF Scan function off
Press “UP” or “DOWN” to toggle the scan on or off.
Sc=On Scan function on
Press “SET” to accept the present setting and continue.
SAccept scan setting
7.3.2 Scan Rate
The next function in the main menu is the scan rate. The scan rate can be set
from .1 to 99.9°C/min. The maximum scan rate howeveris actually limited by
the natural heating or cooling rate of the instrument. This is often less than
100°C/min, especially when cooling.
The scan rate function appears in the main menu after the scan controlfunction.
The scan rate units are in degrees per minute,degrees C or F depending on the
selected units.
Sr= 10.0 Scan rate in°C/min
Press“UP”or“DOWN”tochangethescanrate.
Sr= 2.0 New scan rate
Press “SET” to accept the new scan rate and continue.
SAccept scan rate
9140 Dry-well Calibrator
User’s Guide
24