Alarms included: low inlet pressure, high amp draw, high / low pH

1.5Specifications for E-Series Machines

1.5.1Feed Water Requirements

 

Table 1.1

 

Feed Water Requirements

 

 

Temperature

35° - 77°F (2° - 25°C) Not to exceed

 

85°F (29°C) unless specifically

 

designed for higher temperatures

 

 

Inlet Pressure

Minimum: 30 psig (2.1 barg)

 

Maximum: 60 psig (4.1 barg)

 

 

Chlorine

For Osmo HR(PA) membrane ele-

(continuous feed)

ments: 0 ppm

 

 

Operating pH

Soft water [less than 1 grain per gal-

 

lon: 3.0 - 10.0 (gpg) or 17 mg/L hard-

 

ness], acceptable pH: 5.5 - 6.0

 

 

Pre-filter

5 micron HYTREX cartridge

 

(part number on machine label)

 

 

Inlet Connections

1.5-inch FNPT*

 

 

1.5.2Permeate (Product Water) Flow Rate

Stated on the serial number label (assumes no permeate back pressure, 2000 mg/L TDS maximum feed concentration, and rated temperature).

To estimate permeate output with back pressure, use the formula below:

(Permeate Flow on Label) x [(Operating Pressure) - (Permeate Back Pressure)] Operating Pressure

Permeate Back Pressure:

Maximum: 80 psig (5.5 barg)

Permeate Outlet:

1-inch FNPT

*FNPT: Female National Pipe Thread

7