F-801 Internal RO System by GE Osmonics

1.10Specifications:

1.10.1Input Water Requirements

 

Minimum

Maximum

Input flow Rate

2.0 GPM

N/A

Temperature (water)

10°C (50°F)

32°C (90°F)

NOTE:

Operating at lower temperatures may make it impossible for the F-801 RO to produce adequate permeate for the dialysis machine.

pH (operating)

5.5

10.5

TF Membranes

 

 

pH w/chloramines present (operating)

5.5

8.5

TF Membranes

 

 

pH (cleaning)

2.0

11.0

TF Membranes

 

 

Pressure, dynamic

30 PSI

100 PSI

1.10.2Flow Rates

Product Flow based on the following conditions:

 

New Membranes

 

 

Operating Pressure:

150 PSI

 

Product Pressure:

1 PSI

 

Feed Water Temperature:

77°F (25°C)

GPM

LPM

GPD

LPD

 

.41

1.55

588

2225.8

 

Membrane flux may vary ±15%

Waste Flow: 2000ml/min (±10%)

1.10.3Electrical Requirements

Motor

1/3 HP

 

Voltage

115 V~

 

Hertz

60

 

Amps

10

 

Phase

single

 

Power Cord (length)

8 ft (2.4 m)

 

 

 

 

 

 

 

12552e – 27 Feb 04

1 - 12