1.10.4Membrane Performance Characteristics
Salt rejection (TF Membranes): | 95% minimum at initial testing at factory. |
Test Solution: | Inlet water 1000 PPM NaCl |
Conductivity: | 2000 micro |
Operating Pressure: | 200 PSI |
Product Back pressure: | 0 PSI |
1.10.5Environmental Requirement
| Minimum | Maximum |
Ambient Temperature | 4°C (39°F) | 32°C (90°F) |
Storage Temperature | 2°C (36°F) | 32°C (90°F) |
Altitude | N/A | 10,000 feet |
CAUTION:
This equipment must not be allowed to freeze. Irreparable harm to various components, including RO membranes may result.
NOTE:
The user/operator should recognize that moisture can be caused by condensation and is not necessarily an equipment leak. This equipment will function in the presence of condensation.
1.11Disposal
Disposal of this product or parts must be carried out according with local disposal codes.
1.12Electromagnetic Interference
This equipment can generate, uses and can radiate radio frequency energy and, if not installed and used in accordance with these instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference with other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
•Reorient or relocate the receiving device
•Increase the separation between the equipment
•Connect the equipment into an outlet on a circuit different from that to which the other device(s) is connected
•Consult the field service technician or manufacturer for help.