3.1Daily
WARNING:
Verification of the absence of all sanitizers, cleaners, disinfectants and chlorine/chloramines must be performed before RO product water is used for any patient related uses.
WARNING:
Use of exhausted carbon in the presence of chlorine/chloramines will result in damage to the
CAUTION:
When moving the
CAUTION:
Do not push unit sideways as this may cause the unit to tip over. If it is necessary (such as in tight patient areas), use extreme caution and do not push it in a direction where it could fall onto someone.
CAUTION:
Do not try to catch the equipment if it starts to fall over. If moving up or down a ramp, get help if unable to handle the load.
NOTE:
Obtain a copy of the
1.Connect the RO unit to the required utilities (feed water, drain, and power).
2.Ensure that all the valves in the water supply line to the RO are open.
3.Ensure that all
4.Ensure that the product line is connected to the inlet water supply line of the dialysis machine. Open the product valve located on the end of the product line. If the product valve is closed, no product water will be delivered to the dialysis machine.
5.Turn on the water supply.
12552e – 27 Feb 04 | 3 - 3 | System Operation |