WARNING:
Only the cold water supply should be used. If blending of hot and cold water is to occur, an appropriate temperature blend valve must be used, and the temperature of the incoming water supply must be continuously monitored, and not be allowed to rise above 30°C (86°F). Excessively high water temperature may cause severe patient injury or death.
6.Turn the
CAUTION:
Some pump noise (chattering) is normal in the flush position. Excessive pump noise (chattering) can be damaging to the pump. The RO is in “flush” from the full open (flush) position to just before the fully closed (operate) position. Adjust it by partly closing the flush valve to reduce chattering of the pump.
7.The "Percent Rejection" monitor should indicate 95% rejection within 2 minutes of
8.The TDS monitor should stabilize within 2 minutes of
WARNING:
When the RO is used for dialysis service, a test for disinfectant residual MUST be performed.
9.Perform the following routine checks before each use and record on the Operation Log sheet.
WARNING:
Any data that does not meet minimum standards set at your facility must be reported to the medical director. The RO should not be placed in service until readings are acceptable.
Read the pump pressure gauge, it should read between 140 – 160 PSI.
Read the inlet
Read the product pressure gauge, it should read between 35 – 40 PSI. The product pressure should be read with the product water hose connected to the Fresenius inlet, or with the product outlet valve closed.
12552e – 27 Feb 04 | 3 - 4 | System Operation |