Millenium™ RO System by GE Osmonics
valve allowing product flow. When the hose is disconnected, the product flow is stopped.
4.Make sure that the inlet water valve switch and the RO pump power switch are in the OFF position. Plug the power cord into a 115 VAC (230 VAC) grounded outlet. Do not use an extension cord or an adapter.
5.Turn the water supply to the RO on.
6.Turn the operate/flush valve to the flush position.
7.Turn the inlet water valve switch on. Water will start to fill the RO. Allow the RO to rinse to drain for five minutes after water begins to flow from the drain line. The pump pressure gauge should indicate some pressure.
8.Turn the RO pump power switch on.
NOTE:
The operate/flush valve may need to be rotated part way towards the operate position if the pump makes excessive noise.
9.Slowly turn the operate/flush valve to the operate position. Pump pressure should not exceed 200 PSI. If 200 PSI is exceeded, adjust pump pressure on pump head located inside the cabinet. See instructions in Figure 1 below:
Figure 1
14693e - 31 Mar 03 | 2 - 4 | Installation Guidelines and Instructions |