Millenium™ RO System by GE Osmonics

The operate/flush valve may need to be rotated partially closed, (towards the operate position) if the pump makes excessive noise.

6.Rotate the operate/flush valve to the operate position.

7.Turn the RO pump power switch off for 1 minute. Turn the RO pump switch on for 2-3 minutes. This will aid in the rinsing out of any potentially trapped disinfectant in the Millenium™ RO.

8.Continue to rinse for an additional 5 minutes.

9.Check the product line for residual disinfectant with an appropriate test kit. If the test is positive, continue to rinse until a negative result is achieved.

10.Disconnect the product rinse line and store it.

11.Remove the 'Do Not Use/Contains Disinfectant’ warning label. The Millenium™ RO System is now ready for use.

14693e - 31 Mar 03

4- 25

Routine Maintenance