Millenium™ RO System by GE Osmonics
The operate/flush valve may need to be rotated partially closed, (towards the operate position) if the pump makes excessive noise.
6.Rotate the operate/flush valve to the operate position.
7.Turn the RO pump power switch off for 1 minute. Turn the RO pump switch on for
8.Continue to rinse for an additional 5 minutes.
9.Check the product line for residual disinfectant with an appropriate test kit. If the test is positive, continue to rinse until a negative result is achieved.
10.Disconnect the product rinse line and store it.
11.Remove the 'Do Not Use/Contains Disinfectant’ warning label. The Millenium™ RO System is now ready for use.
14693e - 31 Mar 03 | 4- 25 | Routine Maintenance |