© National Instruments Corporation 25 NI 5421 Specifications
Delay from
Start Trigger to
CH 0 Analog
Output
Interpolation Factor Typical Delay Refer to ts2 at
NI Signal
Generators
Help»Devices»
NI 5421»
NI <bus>-5421»
Triggering»
Trigger Timing.
Digital Interpolation Filter
disabled.
43 Sample Clock
Periods + 110 ns
257 Sample Clock
Periods + 110 ns
463 Sample Clock
Periods + 110 ns
864 Sample Clock
Periods + 110 ns
Delay from
Start Trigger to
Digital Data
Output
40 Sample Clock periods + 110 ns. —
Trigger Exporting
Exported
Trigger
Destinations
A signal used as a trigger can be routed out to any
destination listed in the Destinations specification
of Table9.
—
Exported
Trigger Delay
65 ns (typical). Refer to ts3 at NI Signal Generators Help»
Devices»NI 5421»NI <bus>-5421»Triggering»Trigger
Timing.
—
Exported
Trigger Pulse
Width
>150 ns. Refer to ts4 at NI Signal Generators Help»
Devices»NI 5421»NI <bus>-5421»Triggering»Trigger
Timing.
—
Table 8. (Continued)
Specification Valu e Comments