PROCEDURE for Free Iron
applies to DR/4000, DR/2500 DR/2010 and DR/800
DR /
HCT 159
MEASURING RANGE:
0.10 - 5.00 mg/L Free-Fe / 0.12 - 6.00 mg/L Tot-Fe
1 |
2 |
4.0ml
Sample
1. Enter program number and set up the instrument as explained for each instrument type.
2. Pipet 4.0 ml sample into the sample vial.
Note
If this is the first time you have run this revision of the test on your DR/4000, you must contact your local Hach distributor for an upgrade of your software (Cat.No.
INSTRUMENT SET-UP
1. | Press the soft key under HACH PROGRAM (or |
| USER PROGRAM if the method was entered as a |
| USER PROGRAM). |
| Enter program number 2130 (or 159 if the method was |
| entered as a USER PROGRAM). |
| Press ENTER. |
2. | The display will show: |
| HACH PROGRAM: 2130 |
| Iron, HCT 159 |
| The wavelength ( λ), 485 nm, is automatically selected. |
3. | Insert the Test N’Tube adapter into the cell holder by |
| sliding it under the thumb screw and into the alignment |
| grooves. Fasten in place by tightening the thumb screw. |
4. | Follow the procedure steps. |
| (Note: If measuring Total Iron and performed the sample |
| digestion using the HCT 200 Metal Prep Set, select the |
3
415min
3. Close sample vial and invert a few times, until the
4. Start the instrument timer.
A
| Total Iron alternate form under the Options soft key menu). |
5. | Wipe the outside of the zero vial. |
| Place the zero vial into the adapter. |
| Close the light shield. |
6. | Press the soft key under ZERO. |
| The display will show: 0.00 mg/L |
0.00 mg/L
7.Clean the outside of the sample vial.
8.Place the prepared sample vial into the adapter. Close the light shield. Results in mg/L
Note
If the test overranges, repeat the UniCell™ test with diluted sample.
5 |
ZERO |
5. Zero the instrument using the zero vial (white cap) provided and read the sample.
LIST OF REQUIRED REAGENTS |
|
|
|
| |
UniCell™ HCT 159 | 23/pkg | HCT 159 |
OPTIONAL REAGENTS |
|
|
Metal Prep Set |
|
|
UniCell™ HCT 200 | 50 digestions | HCT 200 |
Iron Standard |
|
|
1000 mg/L as Fe | 100 ml | |
Sodium Hydroxide |
|
|
Standard Solution | 1000 ml | |
(5.0 N) |
|
|
Nitric Acid, concentrated | 500 ml |
PROCEDURE for Total Iron:
After digestion with the Metal Prep Set HCT 200, follow the procedure steps 1 – 5.