Function and Instruction For 2-8 Pharmaceutical
Refrigerator
Control pannel
5 Adjust alarm temperature range: See fig #1
Battery switch
Two methods set high temperature alarm “HA” and low temperature alarm “LA”
Press “Alarm” button for 10 seconds. LED display code “HA”,press “SET” button again(quickly),code
“HA” flash, adjust the set- temperature by both pressing “+” and ” buttons and increment 1°C, adjustment range shouldn't be 6°C higher than
display cabinet temperature) and LED display “HA” Press both “+” and ” to choose another parameter “LA”, “LA” setting method like “HA”, and its adjustment range shouldn't be 6° C lower than
6 Defects display
Inner light
The control panel only can display one sensor error, when the sensor error happens.LED display “E1”,and alarm at the same time. The refrigerator will keep working according to the defect safty model,and you should put the stored goods to other refrigerator as soon as possible,and cantact the
7Inner light
In order to see the stored goods clearly, a LED light has been installed in the cabinet. You can control light “on” or “off”by “light” button to save energe,please turn off the light after observing the stored goods in the refrigerator.
The light is on when the door is open; The light is off when the door is closed.
Door handle
Shelf
Casters