Haier HYC-68/68A Function and Instruction For 2-8 Pharmaceutical Refrigerator, Defects display

Models: HYC-68/68A

1 8
Download 8 pages 59.92 Kb
Page 6
Image 6
Function and Instruction For 2-8  Pharmaceutical

Function and Instruction For 2-8 5 Adjust alarm temperature range: See fig #1 Pharmaceutical

Refrigerator

Control pannel

5 Adjust alarm temperature range: See fig #1

Battery switch

Two methods set high temperature alarm “HA” and low temperature alarm “LA”

Press “Alarm” button for 10 seconds. LED display code “HA”,press “SET” button again(quickly),code

“HA” flash, adjust the set- temperature by both pressing “+” and 6 Defects display7Inner light ” buttons and increment 1°C, adjustment range shouldn't be 6°C higher than pre-set cabinet temperature,press “Alarm” button to return the former menu(each press of “Alarm” button will return one step back to former menu until

display cabinet temperature) and LED display “HA” Press both “+” and Manual backgroundManual background ” to choose another parameter “LA”, “LA” setting method like “HA”, and its adjustment range shouldn't be 6° C lower than pre-set cabinet temperature.

6 Defects display

Manual background Inner light

The control panel only can display one sensor error, when the sensor error happens.LED display “E1”,and alarm at the same time. The refrigerator will keep working according to the defect safty model,and you should put the stored goods to other refrigerator as soon as possible,and cantact the after-sales department quickly.

7Inner light

In order to see the stored goods clearly, a LED light has been installed in the cabinet. You can control light “on” or “off”by “light” button to save energe,please turn off the light after observing the stored goods in the refrigerator.

The light is on when the door is open; The light is off when the door is closed.

Door handle

Shelf

Casters

Page 6
Image 6
Haier HYC-68/68A Function and Instruction For 2-8 Pharmaceutical Refrigerator, Adjust alarm temperature range See fig #1