1.2Installation Requirements
1.2.6Vibrations
Before installing the instrument, measure floor vibration at the site. Image trouble will not occur if the allowable values in Table
Check the vibration conditions at the site before installation (or upon receiving an order) as required. Make sure that the values shown in the table are satisfied.
External vibration: Image trouble will not occur if the conditions in Table
Table
Type I Stage Allowable Value
Frequency | X, | |
2.0 Hz | 0.8 µm | - |
3.0 Hz | 0.8 µm | - |
5.0 Hz | 0.8 µm | 4.0 µm |
8.0 Hz | 2.0 µm | 1.5 µm |
10.0 Hz | 3.0 µm | 2.5 µm |
Type I Stage Allowable Value
Frequency | X, Y, | |
10.0 Hz | 1.0 cm/s2 | max. |
50.0 Hz | 2.0 cm/s2 | max. |
60.0 Hz | 2.0 cm/s2 | max. |
Type II Stage Allowable Value
Frequency | X, | |
2.0 Hz | 1.0 µm | - |
3.0 Hz | 3.0 µm | - |
5.0 Hz | 2.0 µm | 0.6 µm |
8.0 Hz | 1.5 µm | 1.2 µm |
10.0 Hz | 2.5 µm | 1.6 µm |
1 - 10