SC Series Instruction Manual page 131
DAC Output Channel14
14.3 Setup Procedure
If you are not familiar with operating the instrument in the SETUP menu mode,
see “SETUP Menu mode” on page23. A listing of all menu items is given in
“Setup Menu Reference” on page157.
Step 1: Determine if you have a Voltage DAC Output channel or a Current DAC
Output channel.
Check the instrument’s Customer Information Sheet, or see “Identifying the
Output Type” on page135.
Step 2: Wire the readout instrument to the channel’s connector.
See the “Wiring” section earlier in this chapter for details.
Step 3: Select the value (channel and source) used to drive the Analog Output
See the “Channel Menu” on page133 for details.
Step 4: Select the full-scale and zero-scale settings used to drive the Analog Out-
put.
See the “Channel Menu” on page133 for details.