Additions to the Installation Guide

The additions to Installation Guide 34-ST-33-39 that relate to the newly designed meter body and process heads are given in Table 1 of this addendum. Use the information in Table 1 to reference and annotate your User Manual.

Table 1 – Additions to the User Manual

Page # in User

Sub-Section

Description of Change

Manual

 

 

 

 

 

19

3.3 Considerations for ST 3000

The Maximum Working Pressure Rating and the

 

Transmitter

Overpressure Rating has been enhanced for all

 

Table 6 Transmitter

models included in this addendum except for the

 

draft range transmitter.

 

Overpressure Ratings

For more information, refer to Table 3 in this

 

 

 

 

Addendum.

 

 

 

38

4.2 Piping ST3000 Transmitter

In Step 2 and in Step 4, the reference to the “Teflon

 

Table 15 Installing Flange

(white) gasket should be “gasket or O-ring”.

 

 

 

Adapter

In Step 5, do not use the torque specification of 47.5

 

 

to 54 N·m (35 to 40 ft-lb). Instead, use the

 

 

following:

 

 

47,5 Nm +/- 2,4 Nm (35 Lb-Ft +/- 1.8 Lb-Ft).

 

 

 

46

5.1 Wiring Diagrams and

The numbers of dimension drawings for the newly

 

Dimension Drawings

designed models are given in Table 2 in this

 

Dimension drawings-Series 100

addendum.

 

 

 

and Series 900

 

 

 

 

2 of 4

34-ST-99-25 (Addendum to 33-ST-33-39)

10/04