Conduct NT – Analyse Recorded Data
Conductance Catheter Settings
Match the catheter with the actual heart size, exclude individual segments and optimize the dual field setting.
Rho measurement
The Rho cuvette is used to measure blood resistivity from a blood sample. Blood resistivity is required for the calibration of the volume measurements. The Rho cuvette is connected to the connector on the right side of the
Calibration
Calibration of the volume signals from the conductance catheter is automatic. However, the 6 user defined signals connected to the external inputs, Pressure, ECG 2, Aux1, Aux2, Aux3 and Aux4 require initial calibration to enable the software to convert input signal voltages to user units (e.g. mmHg).
Display
Select
Acquire Data
Start data acquisition by clicking ‘Acquire’. An audible signal every 1 s reminds the user that acquisition is in progress. Stop data acquisition by clicking acquire again. Enter a comment at the end of each acquisition run.