15.Adjust the injector by performing one of the following:

Bottom of Floating Ball Reads Greater Than 35 ‐ Less disinfectant needs to be added to the mixture. Loosen the screw on the injector to draw less disinfectant into the solution.

Bottom of Floating Ball Reads Less Than 35 ‐ More disinfectant needs to be added to the mixture. Tighten the screw on the injector to draw more disinfectant into the solution.

16.When the bottom of the floating ball is near 35, turn the black adjustment knob on the flow meter until the bottom of the floating ball is at 35.

17.Place the bucket/pail under the burp fitting to collect the disinfectant. Use a funnel if necessary.

18.Note the number of ounces in the graduated plastic bottle.

19.Let the bucket/pail fill to one gallon.

20.Turn the disinfectant valve to the off position.

21.Remove the disinfectant draw tube from the plastic bottle.

22.Examine the amount of disinfectant remaining in the plastic bottle.

23.Subtract the amount of disinfectant remaining from the amount noted in STEP 18.

24.Perform one of the following:

Approximately ½‐ounce of disinfectant has been used ‐ Proceed to STEP 23.

Much more or much less than ½‐ounce of disinfectant has been used ‐ Repeat STEPS 1‐18.

25.Position the disinfectant draw tube into the disinfectant solution.

26.Disinfect the tub. Refer to the tub owner’s manual or to the Closed Loop Disinfecting Procedure instruction sheet, part number 1154273.

Service Access

Cabinet

Disinfectant Siphon Tube

Injector

Screw

Disinfectant Valve

DETAIL “A” -

FLOW METER

Flow

Adjustment

Meter

 

Knob

Floating

Ball

FIGURE 1 Calibrating the Disinfectant Flowmeter

Invacare Continuing Care Group

www.invacare-ccg.com

 

 

 

USA

Canada

 

One Invacare Way

994 Hargrieve Rd.

All rights reserved. Trademarks are identified by the symbols ™ and ®. All

Elyria, Ohio USA

London Ontario

trademarks are owned by or licensed to Invacare Corporation unless

44036-2125

N6E IP5 Canada

otherwise noted.

 

 

Betadine is a registered trademark of Purdue Pharma L.P.

 

 

© 2008 Invacare Corporation

 

 

Part No 1154272

For Customer Service call: 800-668-2337

Rev A - 04/08

Page 2
Image 2
Invacare 1154272 owner manual Detail a Flow Meter, Floating Ball, Invacare Continuing Care Group, Usa, Canada