ASSEMBLY OF THE EXISTING PUMP, NEW BOOM AND SWIVEL BAR ASSEMBLIES
Pump Assembly
1.Using the nut and bolt provided, assemble the pump to the new boom. Refer to FIGURE 2.
Boom Assembly (FIGURE 4)
1.Align the holes of the boom assembly and mast as- sembly mounting bracket. Insert the bolt completely through the holes of the mast assembly mounting bracket and the boom assembly.
NOTE: Be sure that the bolt is completely through the holes of the mast assembly mounting bracket and boom assembly; and they are not riding on the shoulder of the bolt.
CAUTION
The boom and mast when assembled MUST pivot easily.
2.Tighten the bolt and nut using an Allen wrench and an adjustable wrench.
NOTE: The nut and bolt can be tightened completely and the boom assembly will pivot easily if the mounting hardware is aligned properly when the boom assembly is secured to the mast. To check for correct alignment of the hardware, use an Allen wrench and turn the bolt by hand (it should rotate withoutalotofforcebeingapplied).Ifexcessiveforce is necessary, the bolt is NOT aligned properly. Disassemble and repeat assembly procedures until proper alignment is obtained.
Swivel Bar Attachment with Mounting Bracket (FIGURE 5)
1.Pull the pin up through the swivel bar pad.
2.
NOTE: The swivel bar has hooks on both ends to accept the hardware used to attach the sling that supports the patient during lift. The swivel bar pad should remain in place during normal use.
Nut
Pin
Shoulder
Screw
Swivel Bar
Pad
NOTE: Make sure Pin and Washers are completely pushed UP through the Swivel Bar Pad.
FIGURE 5 - SWIVEL BAR ATTACHMENT
WITH MOUNTING BRACKET
NOTE: Refer to Owner's Manual for proper operation of your patient lift.
FIGURE 4 - NEW BOOM INSTALLATION
899 | Cleveland Street P.O. Box 4028 Elyria, Ohio | ||
Form | No. | Part No. 1060540 | Printed in U.S.A. |