CALIBRATION OF THE RELIANT SCALE
NOTE: The Reliant Scale will be
1.The patient and the sling must be removed from the scale to properly calibrate the Reliant Scale. For removing the patient instructions refer to the Sling Owner's Manual - Part No. 1023891. For sling detachment instructions refer to the Sling Owner's Manual - Part No. 1023891.
2.With the Reliant Scale on, remove the four screws on the back of the enclosure and remove front cover to expose the PC board.
3.Press the CAL button located on the PC board. The CAL switch is not labeled but is the only button on the PC board located in the lower right corner. Once the CAL button is pushed "CAL1" will be seen in the display window.
NOTE: CAL1 selects the calibration mode using 50 pounds of calibrated weight. CAL2 selects the calibration mode for use with 200 pounds of calibrated weight. Pressing the UNITS key toggles between CAL1 and CAL2.
4.When the desired calibration mode is displayed, press the LOCK/UNLOCK key. NOTE: The display window will now show UnLd.
5.Ensure that there is no load on the scale and press the ZERO key.
6.The display window will show a dashed line
7.Perform one (1) of the following:
A.FOR CAL1 OPTION - support 50 pounds of calibrated weight from the scale and press the ZERO key.
B.FOR CAL2 OPTION - support 200 pounds of calibrated weight from the scale and press the ZERO key.
8.The display window will show a dashed line
9.Remove the weight from the scale and press the LOCK/UNLOCK key.
NOTE: The unit will now be in a temporary test mode and will not lock the weight display. This will allow weight to be loaded and unloaded to check the calibration.
10.Turn the unit off by pressing the ON/OFF key.
11.Turn the unit on by pressing the ON/OFF key. NOTE: Scale will now be in normal operation .
Reliant Scale | 8 | Part No. 1114832 Rev B |