RESEARCH MICROSCOPE
APPLICATION
The
This product, with a modern design and
SPECIFICATIONS
•10x Widefield Eyepieces w/ pointer
•30° Inclined Rotating Head
•4x, 10x, 40xR, 100xR DIN Achromat Objective Lens
•Coarse and Fine Coaxial Focal Adjustment
•Seidentopf Binocular Head
•Rack & Pinon Abbe NA 1.25 Condenser
•Low Position Coaxial Mechanical Stage
•20 watt Halogen Lamp
OPERATION
Lamp Illumination
Adjust the brightness control switch to achieve the desired brightness. The degree of light intensity depends on individual conditions such as specimen contrast, objective magnification, individual eyesight, etc. Too little or too much light is not recommended. Usually do not turn the light to the full brightness. This can shorten the life of the bulb.
To center the lamp, loosen the adjustment screw on the bottom of the pull down door, center the image of filament in the view field, then tighten the adjust- ment screw so as to observe at the optimum illumination.
Focus
Place a specimen slide on the center of stage. Using the 4x or 10X objective and 10X eyepiece, raise the stage as high as possible.
Slowly rotate the coarse focus knob to lower the specimen while looking though the 10X eyepieces. Use the fine focus knob to obtain a sharp image when changing to other objectives, the
The tightness of the coarse focus knob has been adjusted when the micro- scope leaves factory. If it is too loose (the stage falls automatically), please adjust the tension using the coarse adjustment tension knob by turning counter- clockwise until it is suitable.
Condenser
Adjust the center and height of the condenser. It has been adjusted at a suit- able position when the microscope leaves factory; the user may not need any adjustment. It can be removed if you unscrew the clamp screw.
The surface of the condenser should be 0.2mm lower than the stage surface when the condenser raises to its full limit. The condenser can be moved up and down by rotating the focus knob of condenser. When using the high power objectives, the condenser should be risen. When using the low power objec- tives, the condenser should be lowered.
Place the blue filter into the holder, then insert into the iris diaphragm. To remove, reverse process.
BULB REPLACEMENT
Before replacing the lamp, turn OFF the power switch and unplug from elec- trical source. Pull lamp door down and remove lamp by pulling straight out of socket.
When replacing the lamp, do not touch the glass of the bulb with bare hands. Wear gloves or keep the protective cover on the bulb during installation. Wipe off any fingerprints using clean cloth.
CARE AND MAINTENANCE
The
Cleaning the Microscope
Dust is best removed with a soft brush or soft cloth. More persistent dirt, such as fingerprints, grease and oil, may be removed with soft cotton or lens tissue, lightly moistened with lens cleaner.
To clean immersion oil off the
When the microscope is not in use, cover it up with dust cover, and store in a dry place not subject to mold.
kav.instrcman.T2701.pc.v1