EMC Guidance

Guidance and manufacture’s declaration - electromagnetic emissions

The Oxygen Saturation Monitor Model PULSOX-300 is intended for use in the electromagnetic environment specified below. The customer or the user of the Model PULSOX-300 should assure that it is used in such anenvironment.

Emissions test

Compliance

Electriomagnetic environment - guidance

RF emissions

 

The Model PULSOX-300 uses RF energy only for

CISPR 11

Group 1

its internal function. Therefore its RF emissions are

 

 

very low and are not likely to cause any interference

 

 

in nearby electromagnetic equipment.

RF emissions

Class B

The Model PULSOX-300 is suitable for use in all

CISPR 11

establishments, including domestic establishments

 

 

 

and those directly connected to the public

Harmonic emissions

 

Not applicable

low-voltage power supply network that supplies

IEC61000-3-2

 

buildings used for domestic purposes.

Voltage fluctuations/

Not applicable

 

flicker emissions

 

IEC61000-3-3

 

 

Appendix

Guidance and manufacture’s declaration - electromagnetic immunity

The Oxygen Saturation Monitor Model PULSOX-300 is intended for use in the elec-tromagnetic environment specified below. The customer or the user of the Model PULSOX-300 should assure that it is used in such anenvironment.

Immunity test

IEC 60601

Compliance

Electromagnetic environment -

 

Test level

level

guidance

 

 

 

 

Electromagnetic

±6 kV contact

±6 kV contact

Floors should be wood, concrete or

Discharge (ESD)

±8 kV air

±8 kV air

ceramic tile, if fllors are covered with

IEC61000-4-2

 

 

synthetic materisl, the relative humidity

 

 

should be at least 30%.

 

 

 

Electrical fast

Not applicable

 

 

Transient/burst

 

 

 

IEC61000-4-4

 

 

 

Surge

Not applicable

 

 

IEC61000-4-5

 

 

 

 

 

 

 

Voltage dips, short

Not applicable

 

 

Interruptions and

 

 

 

Voltage variations

 

 

 

on power supply

 

 

 

Input lines

 

 

 

IEC61000-4-11

 

 

 

 

 

 

 

Power frequency

3 A/m

3 A/m

Power frequency magnetic fields

(50/60Hz) magnetic

 

 

should be at level characteristic of a

field

 

 

typical location in a typical commercial

IEC61000-4-8

 

 

or hospital environment.

 

 

 

 

 

 

 

26