?WHAT DISPLAY SYMBOLS MEAN
Display/ | Condition/Cause | Recommended | ||||||
Symbol | Action | |||||||
| ||||||||
|
|
|
|
|
|
|
| |
|
|
|
|
|
| The symbol shown while measure- | Measurement is in progress, | |
|
|
|
|
|
| ment is in progress. It blinks while | ||
|
|
|
|
|
| remain as still as possible. | ||
|
|
|
|
|
| detecting the pulse. | ||
|
|
|
|
|
| |||
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
| |
|
|
|
|
|
| An irregular heartbeat or body |
| |
|
|
|
|
|
| movement may have occurred. | Take measurement again and | |
|
|
|
|
|
| Refer to page | ||
|
|
|
|
|
| consult with your physician. | ||
|
|
|
|
|
| mation on irregular heartbeats. | ||
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
| |
|
|
|
|
|
| Previous measurement stored in |
| |
|
|
|
|
|
| MEMORY. |
| |
|
|
|
|
|
| The battery power indicator during | Replace all batteries with new | |
|
|
|
|
|
| |||
|
|
|
|
|
| measurement. | ||
| Battery Full | |||||||
|
|
|
|
|
|
| ones when the symbol blinks. | |
|
|
|
|
|
| The batteries are low when it | ||
|
|
|
|
|
| Alkaline batteries recommended. | ||
|
|
|
|
|
| blinks. | ||
| Battery Low |
| ||||||
|
|
|
|
|
| Unstable blood pressure due to | Try the measurement again. | |
|
|
|
|
|
| excessive body movement. | Remain very still during the | |
|
|
|
|
|
|
| measurement. | |
|
|
|
|
|
| The systolic and diastolic values are | Fasten the cuff correctly, and try | |
|
|
|
|
|
| within 10 mmHg of each other. | ||
|
|
|
|
|
| the measurement again. | ||
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
| |
|
|
|
|
|
| The pressure value did not increase | Check for air leaks along the | |
|
|
|
|
|
| during inflation. | tube and around the air socket. | |
|
|
|
|
|
| The cuff is not fastened correctly. | Refasten the cuff and retake mea- | |
|
|
|
|
|
|
| surement. | |
|
|
|
|
|
| There is an air leak in the cuff or | Make sure tube is properly con- | |
|
|
|
|
|
| monitor. | nected to cuff and monitor. | |
|
|
|
|
|
|
| Try the measurement again. | |
|
|
|
|
|
| The pulse is not detected correctly. | Remain very still during the | |
PULSE DISPLAY ERROR |
| measurement. | ||||||
|
|
|
|
|
|
|
| |
|
|
|
|
|
| Cuff Inflation Meter | Measurement is in progress, | |
|
|
|
|
|
| remain as still as possible. | ||
|
|
|
|
|
| |||
|
|
|
|
|
|
|
| |
|
|
|
|
|
| Pressure Rating Indicator™ | Refer to the section “About | |
|
|
|
|
|
| |||
|
|
|
|
|
| Pressure Rating Indicator” for | ||
|
|
|
|
|
| |||
|
|
|
|
|
|
| further explanation. | |
|
|
|
|
|
|
|
| |
|
|
|
|
|
| The alarm is turned on. | Press AM button once to turn | |
|
|
|
|
|
| alarm off. | ||
|
|
|
|
|
|
|
LifeSource Health Line |