MODEL 306 DUAL CHANNEL DATALOGGING THERMOMETER
1.0 Introduction:
This instrument is a dual channel, recording digital thermometer for use with any Type K thermocouple as a temperature sensor. Temperature indication complies with the NIST and IEC584 temperature/voltage tables for Type K thermocouples. Its internal memory can store up to 16312 records in any number of recording intervals. It is equipped with a RS232 interface for
2.0 Specifications:
Measurement Range:
Accuracy: (At ambient of 23 ± 5°C)
MODEL 306 DUAL CHANNEL DATALOGGING THERMOMETER
5. | Exit: |
| Terminates ThermoLog program. |
6. | Help: |
| |
7. | DataLogger: |
| The DataLogger window controls the loading of recorded data from thermometer. |
8. | Tabular: |
| The Tabular window displays the present data from the thermometer in a scrolling table. |
| Data can be stored as a file, or the table can be copied to other software, such as EXCEL, |
| for further analysis. |
9. | Tray Icon: |
Range | Accuracy |
±(0.2% reading + 1°C) | |
200°C - 400°C | ±(0.5% reading + 1°C) |
400°C - 1370°C | ±(0.2% reading + 1°C) |
±(0.5% reading + 2°F) | |
±(0.2% reading + 2°F) | |
200°F - 2498°F | ±(0.3% reading + 2°F) |
Note:
Temperature Coefficient:
For ambient temperatures from 0°C - 18°C and 28°C - 50°C, for each °C ambient below 18°C or above 28°C, add the following tolerance into the accuracy spec.
0.01% of reading + 0.03°C
(0.01% of reading + 0.06°F)
When ThermoLog is running, an icon appears in the Windows Tray area (see figure below), |
to display a |
The basic accuracy specification does not include the error of the probe. Please refer to the probe accuracy specification for additional details.
Resolution: | 200°C | 1°C | |
| outside this range: | 1°F | |
Number of Sensor Inputs: Two Type K Thermocouples |
| ||
Input Protection at Thermocouple Input: 60V DC or 24Vrms AC |
| ||
Display: | Three 4 Digit Temperature Displays plus Indicators, see Section 3.0 | ||
| Switchable backlight |
|
|
Sample Rate: | 1.25 times per second |
|
|
Time Function: Real time clock, see Section 4.6
Datalogging Capacity: 16,312 records
Recording Interval: User adjustable, see Section 4.7
Digital Output:
Power requirement: 9 Volt Battery, NEDA 1604 or JIS 006P or IEC6F22
Battery Life: Approx. 100 hours with alkaline battery; Low battery indication
Operating Conditions:
ξ Operating Temperature and Humidity: 0°C - 50°C (32°F - 122°F); 0 - 80% RH
ξ Storage Temperature and Humidity:
ξ Altitude: Up to 2000 meters (6500 feet). Dimensions: 184×64×30 mm (7.3×2.5×1.2 in)
Weight: Approx. 210g (7.4 oz)
Accessories: Two Type K Wire Thermocouple Probes, Battery, Carrying Case, Instruction Manual, Software, RS232 Cable
Option : AC Adapter: 9VDC ±15%; 100mA: Plug Diameter: 3.5mm×1.35mm
1