Chapter 2 Function Reference — AO_Change_Para meter
NI-DAQ FRM for PC Compatibles 2-32
©
National Instruments Corporation
Output Enable
On some of the devices, you can disable the output even when the waveform generation is in
progress. You can use this feature to bring the output to a known level at any time.
On DAQArb 5411 devices, there is a relay just in front of the I/O connector. By disabling th e
output, this relay switches so that the I/O connector output shorts to ground. The waveform
generation can still continue, but no signal appears at the I/O connector output. You can
enable or disable the output at any time.
To change the output enable setting, set paramID to ND_OUTPUT_ENABLE.
Output Impedance
On some of the devices, you can select the output impedance to match the impedance of
theload.
Output impedance of 50 is good for testing most of the devices. You can use output
impedance of 75 for video testing. If you select an output impedance of 0 , you should be
driving an unterminated load (that is, a load with a very high input impedance).
To change the output impedance setting, set paramID to ND_IMPEDANCE.
Note The values are set up in milliOhms (m).
Table 2-9. Parameter Setting Information for Output Enable
Device Type Per Channel
Selection Possible Legal Range
for paramValue Default Setting
for paramValue
DAQArb AT-5411
DAQArb PCI-5411
4451 devices
4551 devices
Yes ND_YES and
ND_NO ND_NO
Table 2-10. Parameter Setting Information for Output Impedance
Device Type Per Channel
Selection Possible Legal Range
for paramValue Default Setting
for paramValue
DAQArb AT-5411
DAQArb PCI-5411 Yes 0, 50,000, and
75,000 50,000