Chapter 2 Hardware Overview
NI 5911 User Manual 2-10
©
National Instruments Corporation
Figure 2-4. Below-Level Analog Triggering Mode
In below-level analog triggering mode, the trigger is generated when the
signal value is less than triggerValue. hysteresisValue is unused.
Figure 2-5. Above-Level Analog Triggering Mode
In above-level analog triggering mode, the trigger is generated when the
signal value is greater than triggerValue. hysteresisValue is unused.
Trigger Value
Falling Edge Trigger
Rising Edge Trigger
Trigger Value
Falling Edge Trigger
Rising Edge Trigger
CBIHWum.book Page 10 Thursday, October 29, 1998 1:56 PM