NI 9219 Operating Instructions 10 ni.com
Wiring TEDS ChannelsThe NI 9219 supports only Class II
TEDS sensors. Connect the two
TEDS lines to
TEDS Data (T+) and TEDS COM (T–) and ensure
that neither T+ nor T– is tied in common to any of the signal inputs
(terminals 3 through 6) on the NI 9219. Visit ni.com/info and
enter the info code rdteds for information about TEDS sensors.
Grounding and Shielding ConsiderationsYou can connect ground-referenced or floating signal sources to
the NI 9219. If you make a floating connection between the signal
source and the NI 9219, make sure the voltages on the positive and
negative connections are within the channel-to-earth working
voltage range to ensure proper operation of the NI 9219. Refer to
the Specifications section for more information about operating
voltages and overvoltage protection.
Note For best signal quality, National Instruments
recommends using shielded cables and twisted pair
wiring whenever possible.
Figures3 and 4 illustrate connecting grounded and floating signal
sources to the NI 9219 in Voltage mode.