NI 9264 Operating Instructions and Specifications 14 ni.com
Protection
Overvoltage................................ ±27 V at 25 °C
Short-circuit................................Indefinitely
Update time
Noise.................................................500 μVrms
Crosstalk............................... ............ 90 dB
Settling time (100 pF load, to 1 LSB)
20 V step.....................................20 μs
1V step.. .....................................15 μs
0.1V step ....................................13 μs
Number of
Channels
Update Time for
cRIO-9151 R Series
Expansion Chassis
Update Time for All
Other Chassis
13.7μs min 3.1μs min
26.6μs min 5.3μs min
39.4μs min 7.5μs min
16 47 μs min 37 μs min