© National Instruments Corp. 13 NI 9265 Operating Instructions and Specifications
Full-scale output current
Minimum.................................... 20.1 mA
Typical........................................ 20.6 mA
Maximum.................................. .21.0 mA
Output range........................... ..........0 to 20 mA
Compliance voltage.......................... 12 VDC max
Maximum load..................................600 Ω
Accuracy
Measurement Conditions
Percent of
Reading
(GainError)
Percent of
Range*
(OffsetError)
Calibrated, typ (25°C, ±5 °C) 0.11% 0.19%
Calibrated, max(– 40 to 70 °C) 0.25% 0.4%
Uncalibrated, typ (25°C, ±5 °C) 0.35% 1.4%
Uncalibrated, max(– 40 to 70 °C) 0.85% 2.5%
* Range equals 0 to 20.6 mA