© National Instruments Corp. 5 NI 9425 Operating Instructions and Specifications
Caution For Zone 2 applications, install the system 
in an enclosure rated to at least IP54 as defined by 
IEC60529  and EN 60529.
Caution For Zone 2 applications, connected signals must 
be within the following limit:
Capacitance.......................... 0.03 μF max 
Special Conditions for Hazardous Locations Use in EuropeThis equipment has been evaluated as EEx nC IIC T4 equipment 
under DEMKO Certificate No.03 ATEX 0324020X. Each module 
is marked   II 3G and is suitable for use in Zone 2 hazardous 
locations. If you are using the NI9425 in Gas Group IIC hazardous 
locations or in ambient temperatures of –40°C ≤Ta ≤70 °C, you 
must use the device in an NI chassis that has been evaluated as 
EExnC IIC T4, Ex nA IIC T4, or Ex nL IIC T4 equipment.