© National Instruments Corp. 19 NI 9485 Operating Instructions
Shock and Vibration
To meet these specifications, you must panel mount the system
and either use ferrules to terminate wires to the detachable
screw-terminal connector or use the NI9939 backshell kit to
protect the connections.
Operating vibration
Random (IEC 60068-2-34).........5 grms, 10 to 500 Hz
Sinusoidal (IEC 60068-2-6)....... 5 g, 10 to 500 Hz
Operating shock
(IEC 60068-2-27)..............................30 g, 11 ms half sine,
50 g, 3 ms half sine,
18 shocks at 6 orientations
Electromagnetic Compatibility
This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement,
control, and laboratory use:
EN 61326 EMC requirements; Industrial Immunity
EN 55011 Emissions; Group 1, Class A
CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A