NI PXI/PCI-5114 Specifications 16 ni.com
Environment

NI PXI-5114

Note To ensure that the NI PXI-5114 cools effectively, follow the guidelines in
the Maintain Forced Air Cooling Note to Users included in the NI PXI-5114 kit.
TheN I PXI-5114 is intended for indoor use only.
Specification Value
Operating Temperature 0 ºC to +55 ºC in all NI PXI chassis except the following:
0 ºC to +45 ºC when installed in an NI PXI-1000/B or PXI-101x chassis
Meets IEC-60068-2-1 and IEC-60068-2-2
Storage Temperature –40 ºC to +71 ºC
Meets IEC-60068-2-1 and IEC-60068-2-2
Operating Relative
Humidity
10% to 90%, noncondensing
Meets IEC-60068-2-56
Storage Relative
Humidity
5% to 95%, noncondensing
Meets IEC-60068-2-56
Operating Shock 30 g, half-sine, 11 ms pulse
Meets IEC-60068-2-27
Test profile developed in accordance with MIL-PRF-28800F
Storage Shock 50 g, half-sine, 11 ms pulse
Meets IEC-60068-2-27
Test profile developed in accordance with MIL-PRF-28800F
Operating Vibration 5Hz to 500Hz, 0.31 grms
Meets IEC-60068-2-64
Storage Vibration 5Hz to 500Hz, 2.46 grms
Meets IEC-60068-2-64
Testprofile exceeds requirements of MIL-PRF-28800F, Class B
Altitude 2,000m maximum (at 25 °C ambient temperature)
Pollution Degree 2