© National Instruments Corporation A-1 NI PCI-5911 User Manual
ASpecifications
This appendix lists the specifications of the NI5911. These specifications
are typical at 25 °C unless otherwise stated.
Acquisition System
Bandwidth..............................................100 MHz maximum;
refer to Table 2-2, Available
Sampling Rates and
Corresponding Bandwidth in
Flexible Resolution Mode
Number of channels............................... 1
Number of flexible resolution ADC.......1
Max RIS sample rate..............................1 GS/s
Max real-time sample rate......................100 MS/s
Resolution
Sample Rate Mode Effective Resolution
100/n* MS/s Conventional 8 bits
12.5 MS/s Flexible Resolution 11 bits
5 MS/s Flexible Resolution 14 bits
2.5 MS/s Flexible Resolution 15.5 bits
1 MS/s Flexible Resolution 17.5 bits
500 kS/s Flexible Resolution 18 bits
200 kS/s Flexible Resolution 18.5 bits
100 kS/s Flexible Resolution 19 bits
50 kS/s Flexible Resolution 19.5 bits