Chapter 2 Hardware Overview
NI PCI-5911 User Manual 2-12 ni.com
PFI Lines as InputsYou can select PFI 1 or PFI 2 as inputs for a trigger or a reference clock. 
Refer to the Synchronization section for more information about the use of 
reference clocks in the NI 5911.
PFI Lines as OutputsYou can select PFI 1 or PFI 2 to output several digital signals.
Reference Clock is a 10 MHz clock that is synchronous to the 100 MHz 
sample clock on the NI 5911. You can use the Reference Clock to 
synchronize to another NI 5911 configured as a slave device or to other 
equipment that can accept a 10 MHz reference.
Frequency Output is a 1kHz digit al pulse train signal with a 50% duty 
cycle. The most common application of Frequency Output for the NI 5911 
is to provide a signal for compensating a passive probe. 
SynchronizationThe NI5 911 uses a digital phase-locked loop to synchronize the 100 MHz 
sample clock to a 10MHz reference. This reference frequency can be 
supplied by an internal crystal oscillator or through an external frequency 
input through the RTSI bus clock line or a PFI input. 
The NI5911 can also output its 10 MHz reference on the RTSI bus clock 
line or a PFI line so that additional NI5911 devices or other equipment can 
be synchronized to the same reference. 
While the reference clock input is sufficient to synchronize the 100MHz 
sample clocks, it is also necessary to synchronize clock dividers on each 
NI5 911 so that internal clock divisors are synchronized on each device. 
These lower frequencies are important because they are used to determine 
trigger times and sample position.
To synchronize the NI5911 clock dividers, you must connect the digitizers 
with an RTSI bus cable. One of the RTSI bus triggers must be designated 
as a synchronization line. This line is an output from the master device and 
an input on the slave device. To synchronize the digitizers, a single pulse is 
sent from the master NI 5911 to the slaves. This pulse supplies the slave 
devices with a reference time to clear their clock dividers. Hardware arming 
cannot be used during an acquisition using multiple devices. 
To synchronize the triggers of multiple NI5911 devices, one digitizer must 
receive a trigger as described in the Triggering and Arming section and then