© National Instruments Corporation 3 SCC-TC Series Thermocouple Input Modules User Guide
Documentation
One of the following:
SC-2345/2350 User Manual
SCC-68 User Guide
SCC-TC Series Thermocouple Input Modules User Guide
Read Me First: Safety and Radio-Frequency Interference
SCC Quick Start Guide
Tools
1/8 in. flathead screwdriver
Numbers 1 and 2 Phillips screwdrivers
Wire insulation strippers
You can download needed documents from ni.com/manuals.
Note Software scaling of measurements is not supported on the Macintosh operating
system. Refer to the Scaling Voltage Measurements section for more information about
measurement scaling.
Device Specific Information
Note For general SCC module installation and signal connection information, and
information about the SC-2345 or SC-2350 carriers, refer to the SCC Quick Start Guide.

Installing the Module

Caution Refer to the Read Me First: Safety and Radio-Frequency Interference document
before removing equipment covers or connecting/disconnecting any signal wires.
You can plug the SCC-TC0X into any analog input socket on the SCC
carrier. The socket you select determines which E/MSeries DAQ device
channels receive the SCC-TC0X signals.
For single-stage input conditioning, plug the SCC-TC0X into any socket
J(X +1), where X is 0 to 7, and connect the input signals to the module as
described in the Connecting the Input Signals section.
If you use the SCC-TC0X in a dual-stage configuration, the SCC-TC0X
must be the first-stage module. Plug it into any socket J(X+9) and plug the
second-stage SCC into socket J(X+1), where X is 0 to 7. Connect the input
signals to SCC-TC0X as described in the Connecting the Input Signals
section. The SC-2345 connects the output signals of the first-stage SCC to