Chapter 2 Configuration and Installation
SCXI-1161 User Manual 2-8
©
National Instruments Corporation
Jumpers W3, W4, and W5You can use two types of data acquisition boards with the
SCXI-1161-DIO-type boards and MIO-type boards. All three jumpers
have labels indicating the DIO and the MIO configurations. If you want
to use the SCXI-1161 with a DIO-type board, place all three jumpers in
the DIO position, the factory setting. If you want to use the SCXI-1161
with an MIO-type board, place all three jumpers in the MIO position. If
nothing is cabled to the SCXI-1161 rear signal connector, the positions
of these jumpers are irrelevant.
Table 2-2. Configuration of Jumpers W3, W4, and W5 for DIO-Type
andMIO-Type Boards
Jumper Configuration for DIO-Type
Board (Factory Setting) Configuration for
MIO-Type Board
W3
W4
W5
MIO
DIOW3
MIO
DIOW3
MIO
DIOW4
MIO
DIOW4
MIO
DIOW5
MIO
DIOW5