© National Instruments Corporation 13 SCXI-1321 Terminal Block Installation Guide
z is the module slot of the SCXI-1121
w is the channel of the module that you want to engage the shunt and
take measurements
For example, if you want to measure the voltage at channel 0 with the shunt
resistor enabled, use the SCXI channel string
ob0 ! sc1 ! md1 ! shunt0.
You also can specify a list of channels by using
ob0 ! sc1 ! md1 ! shunt0:
w
, for example. Refer to the LabVIEW
Measurements Manual for information on using SCXI channel strings.
The shunting resistors RSCAL are socketed so that you can replace them with
a resistor of another value to achieve the required nulling range for your
application. The sockets and corresponding channels are shown in Table 3.
The factory installed RSCAL provided on the terminal block have a 301 k
±1% value.
Assuming a quarter-bridge strain-gauge configuration with a gauge factor
of GF = 2, the equivalent strain change introduced by the RSCAL shunting
resistor is –199 µε.
Refer to the Traditional NI-DAQ User Manual for more information on
strain-gauge bridge configurations and formulas.
Use the following formula to determine the change due to this shunting
resistor:
Next, using the appropriate strain-gauge strain formula, and assuming that
you have no static voltage, determine the equivalent strain the RSCAL should
provide. For example, RSCAL = 301 k and a quarter-bridge 120 strain
Table 3. Socket to Channel Relationship
Channel Shunt Resistor Socket
0R4
1R6
2R8
3R10
Vchange
VexRdRSCAL Rg
+()
RgRSCAL RdRSCAL Rg
+()+
----------------------------------------------------------------- Vex
2
-------=