© National Instruments Corporation 11 USB-9201/9221 User Guide and Specifications
Sample rate (aggregate)1
USB-9201
Single channel..........................800 kS/s max
Scanning.................................. 500 kS/s
USB-9221....................................... 800 kS/s max
Operating voltage ranges2
USB-9201....................................... ±10 V (not software selectable)
USB-9221....................................... ± 60 V (not software selectable)
Maximum voltage, (AI or COM to earth ground, verified by a dielectric
withstand test)
Screw terminal................................ 250 Vrms
DSUB..............................................±60 VDC
Isolation voltage (AI or COM to earth ground, verified by a dielectric
withstand test)
Screw terminal................................ 2,300 Vrms
DSUB..............................................1,000 Vrms
Overvoltage protection
(AI to COM)...........................................±100 V
Accuracy3
1 Full performance requires the use of a USB 2.0 high-speed host controller and USB 2.0 hubs.
2 Refer to the Safety section for details about safe operating voltages.
3 Excludes noise.
Error
Percent of
Reading
Percent of
Range*
USB-9201
Calibrated typ (25°C, ±5°C) ±0.04% ±0.07%
Calibrated max (0 to 60 °C) ±0.25% ±0.25%
Uncalibrated typ (25 °C, ±5°C) ±0.26% ±0.46%
Uncalibrated max (0 to 60 °C) ±0.67% ±1.25%
USB-9221
Calibrated typ (25°C, ±5°C) ±0.04% ±0.07%
Calibrated max (0 to 60 °C) ±0.25% ±0.25%