Olympus manual ISO9001, ISO14001, 393.1, 380.5, 194.5, 473.7, 473.2, CKX41/CKX31 specifications

Models: CKX41

1 7
Download 7 pages 48.26 Kb
Page 7
Image 7
CKX41/CKX31 specifications

CKX41/CKX31 specifications

Item

 

CKX41

CKX31

Optical system

UIS (Universal Infinity-corrected) optical system

 

Focus

 

Vertical nosepiece movement (stage is fixed),

 

 

 

 

coaxial coarse and fine focus with tension adjustment mechanism, roller slide mechanism,

 

 

 

stroke: 7mm up and 2mm down from focus position which is 1mm above the stage,

 

 

 

stroke per rotation: 39.6mm (coarse), 0.2mm (fine)

 

Revolving nosepiece

Quadruple

 

Stage

 

Plain stage

160mm(L) x 250mm(W), stage clips attachable

 

 

 

 

Exchangeable insert plate (ø25mm opening) incorporated

ø35 petri dish holder stage

 

 

 

 

incorporated

 

 

Mechanical stage

Right-hand low drive coaxial controls,

 

 

 

 

stage movement: X=120mm, Y=78mm, with three dish/sample holders

 

 

 

Substage

70(L) X180(W)mm

 

Illumination

 

Light source

6V,30W halogen lamp, lamp socket (U-LS30-3-2),

 

system

 

 

built-in frosted and heat absorbing filters, detachable illuminator

 

 

 

Filter holder

Insert up to 11mm thick with ø45mm filter, detachable

 

 

 

Aperture diaphragm

Lever type, range: from minimum ø3mm to maximum ø44mm

 

 

 

Slider insertion

With phase slider pocket and built-in slider position click stop mechanism

Condenser

 

Detachable ultra-long working distance condenser (N.A. 0.3, W.D. 72mm)

Contrast slider

Pre-centered phase-contrast: 4X, 10X/20X/40X, empty slot

 

 

 

 

Centerable phase-contrast: 4X, 10X/20X, empty slot (40X optional, pre-centered)

 

 

 

Centerable for Relief Contrast: 10X, 20X, 40X

 

Observation

 

Binocular tube

U-CBI30-2: inclined 30°, interpupillary distance range: 48-75mm,

Fixed binocular tube, inclined

tube

 

 

diopter adjustment by helicoid on left sleeve (F.N. 20)

45°, interpupillary distance range:

 

 

 

U-BI30-2: inclined 30°, interpupillary distance range: 48-75mm,

48-75mm, diopter adjustment by

 

 

 

helicoid on right sleeve

 

 

 

diopter adjustment by helicoid on left sleeve (F.N. 22)

 

 

 

 

 

 

Trinocular tube

U-CTR30-2: inclined 30°, ring dovetail attachment,

 

 

 

 

interpupillary distance range: 48-75mm, tube length

 

 

 

 

and diopter adjustment by helicoid on left sleeve

 

 

 

 

Observation optical path: 50(binocular)/50(video port) (F.N. 20)

 

 

 

 

U-TR30-2: inclined 30°, ring dovetail attachment,

 

 

 

 

interpupillary distance range: 48-75mm, tube length

 

 

 

 

and diopter adjustment by helicoid on left sleeve

 

 

 

Observation optical path: 50(binocular)/50(video port) (F.N. 22)

 

 

Tilting binocular

CKX-TBI: variable inclination angles from 30° to 60°,

 

 

 

tube

interpupillary distance range: 50-76mm,

 

 

 

 

diopter adjustment by helicoid on right sleeve (F.N. 20)

 

 

 

 

UTBI-3: variable inclination angles from 5° to 35°, circular

 

 

 

 

mounting dovetail attachment, interpupillary distance range:

 

 

 

 

50-76mm, diopter adjustment by helicoid on right sleeve. (F.N. 22)

 

Fluorescent illuminator

Detachable illuminator,

 

 

 

 

switchable slide (3-position: B excitation, G excitation, empty slot or U excitation)

 

 

FL light source

50W Hg

 

 

 

FL light shutter

Available

 

 

 

FL field stop

Available

 

 

 

FL cubes

2 cubes (B & G), option U

 

 

 

 

(cubes are not compartible with UIS. Filter and dichroic mirror size are same as UIS)

 

 

Filter

1 filter

 

Eyepiece

 

 

For U-CBI30-2/U-CTR30-2/CKX-TBI: WHB10X/WHB10X-H (F.N. 20)

10X (F.N.: 20)

 

 

 

For U-BI30-2/U-TR30-2/U/TBI3: WH10X/WH10X-H/35WH10X/

 

 

 

 

PWH10X (F.N. 22)

 

Power supply

 

Continuous intensity adjustment,

 

 

 

 

built-in voltage changeover switch (100/120V, 220/240V), frequency 50/60Hz

Objective lens specifications

 

Objective

N.A.

W.D.(mm)

Remarks

For

PLC4X

0.10

22.0

 

brightfield

PLC10X

0.25

10.5

 

ACH20X

0.40

3.0

 

 

LCPLFL40X

0.60

2.6

 

 

UPLFL4XPH

0.13

17.0

PHL (for use with IX2-SL)

 

CPL10XPH

0.25

9.8

PHC (for use with IX2-SL)

For

PL10XPH

0.25

10.5

PH1 (for use with IX2-SL)

LCACH20XPH

0.40

3.2

PHC (for use with IX2-SL)

phase-contrast

LCPLFL20XPH

0.40

6.9

PH1 (for use with IX2-SL)

 

 

LCACH40X2PH

0.55

2.2

PH2 (for use with IX2-SL)

 

LCPLFL40XPH

0.6

2.6

PH2 (for use with IX2-SL)

 

UPLFL4XPHP*

0.13

17.0

For use with IX2-SLP

 

CACH10XPHP*

0.25

8.8

For use with IX2-SLP

 

LCACH20XPHP*

0.40

3.2

For use with IX2-SLP

 

LCACH40XPHP*

0.55

2.2

For use with IX2-SLP

For

CPL10XRC

0.25

9.5

For use with CKX-RC

LCACH20XRC

0.4

2.8

For use with CKX-RC

RC

LCACH40XRC

0.55

1.9

For use with CKX-RC

 

For

UPLFL4X

0.13

17

U,B,G

UPLFL10X

0.3

10

U,B,G

FL

UPLFL20X

0.5

1.6

U,B,G

 

 

LCPLFL20X

0.4

6.9

B,G

 

LUCPLFL40X

0.6

2.7 - 4.0

 

 

 

 

 

 

* Pre-centering objective

CKX41/CKX31 dimensions

 

 

 

(unit: mm)

393.1

473.7

 

380.5

473.2

195

 

 

194.5

 

364

 

 

364

 

370.5

 

237.6

370.5

237.6

515.5

 

486.6

 

 

ISO9001

ISO14001

 

 

Certification

Certification

 

 

Design and production

 

Design and production at the Olympus

 

 

ISO 9001

Accredited by the

adheres to ISO9001

Optical Co. Ltd. Ina Plant conforms with

Accreditation

ISO14001 specifications for

 

Dutch Council for

international quality standard.

 

 

environmental management systems.

Specifications are subject to change without any obligation on the part of the manufacturer.

Page 7
Image 7
Olympus ISO9001, ISO14001, 393.1, 380.5, 194.5, 473.7, 473.2, CKX41/CKX31 specifications, Objective lens specifications