8.Troubleshooting

1)When Calibrating the equipment, the function is not complete

Verify for humidity presence at probes header or at connection box of the equipment.

Another possibility could be the equipments interconnection cable and cell being cut or in short.

2)Reading above Range

First clean the cell and try to calibrate it. If this operation is not succefull, verify the point where the cell is connected, making sure bubbles are not present and if the solution is flowing normally.